Fast cycle, wide temperature range a.c. susceptometer

An a.c. susceptometer of simple design, operating from 4 K to room temperature is described. The sample can be easily replaced and cooled to 4 K in less than 1 h with low helium boil-off. Measuring coils are immersed in the cryogenic bath, so that a serious source of thermal drift voltage is avoided...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Ferrari, H., Codner, G., Melgarejo, D.R., Levy, P., Acha, C., Bekeris, V.
Formato: JOUR
Materias:
Acceso en línea:http://hdl.handle.net/20.500.12110/paper_00112275_v33_n11_p1037_Ferrari
Aporte de:
Descripción
Sumario:An a.c. susceptometer of simple design, operating from 4 K to room temperature is described. The sample can be easily replaced and cooled to 4 K in less than 1 h with low helium boil-off. Measuring coils are immersed in the cryogenic bath, so that a serious source of thermal drift voltage is avoided. © 1993.