Adjustment tolerances for interferometers by amplitude division illuminated with partially coherent light

In this work a general expression is deduced for the mutual intensity function for amplitude division interferometers illuminated with partially coherent sources, where the optical path difference in the branches of the interferometer is taken into consideration. This expression is applied to the pa...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Iemmi, C.C., Simon, J.M.
Formato: JOUR
Acceso en línea:http://hdl.handle.net/20.500.12110/paper_09500340_v38_n5_p847_Iemmi
Aporte de:
Descripción
Sumario:In this work a general expression is deduced for the mutual intensity function for amplitude division interferometers illuminated with partially coherent sources, where the optical path difference in the branches of the interferometer is taken into consideration. This expression is applied to the particular case of Gaussian sources, and an equation for the tolerances for misadjustment of the interferometer is obtained. The parallel between an amplitude division interferometer and an equivalent optical system, analysed in a previous work, is extended to this example. © 1991 Taylor & Francis Group, LLC.