Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions
The images obtained with a classical Czerny-Turner design and a crossed beam Czerny-Turner design are compared, imposing in both cases the Fastie and the Pribram and Penchina conditions. The results show that the astigmatic line may be made parallel to the spectral line only with the classical desig...
Guardado en:
Autores principales: | , , |
---|---|
Formato: | JOUR |
Acceso en línea: | http://hdl.handle.net/20.500.12110/paper_1559128X_v25_n20_p3715_Simon |
Aporte de: |
id |
todo:paper_1559128X_v25_n20_p3715_Simon |
---|---|
record_format |
dspace |
spelling |
todo:paper_1559128X_v25_n20_p3715_Simon2023-10-03T16:25:52Z Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions Simon, J.M. Gil, M.A. Fantino, A.N. The images obtained with a classical Czerny-Turner design and a crossed beam Czerny-Turner design are compared, imposing in both cases the Fastie and the Pribram and Penchina conditions. The results show that the astigmatic line may be made parallel to the spectral line only with the classical design, which makes it superior to the crossed beam design. Also, the importance of an adequate choice of the construction parameters is shown. The analytical calculations are checked by ray tracing. © 1986 Optical Society of America. Fil:Simon, J.M. Universidad de Buenos Aires. Facultad de Ciencias Exactas y Naturales; Argentina. Fil:Gil, M.A. Universidad de Buenos Aires. Facultad de Ciencias Exactas y Naturales; Argentina. Fil:Fantino, A.N. Universidad de Buenos Aires. Facultad de Ciencias Exactas y Naturales; Argentina. JOUR info:eu-repo/semantics/openAccess http://creativecommons.org/licenses/by/2.5/ar http://hdl.handle.net/20.500.12110/paper_1559128X_v25_n20_p3715_Simon |
institution |
Universidad de Buenos Aires |
institution_str |
I-28 |
repository_str |
R-134 |
collection |
Biblioteca Digital - Facultad de Ciencias Exactas y Naturales (UBA) |
description |
The images obtained with a classical Czerny-Turner design and a crossed beam Czerny-Turner design are compared, imposing in both cases the Fastie and the Pribram and Penchina conditions. The results show that the astigmatic line may be made parallel to the spectral line only with the classical design, which makes it superior to the crossed beam design. Also, the importance of an adequate choice of the construction parameters is shown. The analytical calculations are checked by ray tracing. © 1986 Optical Society of America. |
format |
JOUR |
author |
Simon, J.M. Gil, M.A. Fantino, A.N. |
spellingShingle |
Simon, J.M. Gil, M.A. Fantino, A.N. Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions |
author_facet |
Simon, J.M. Gil, M.A. Fantino, A.N. |
author_sort |
Simon, J.M. |
title |
Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions |
title_short |
Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions |
title_full |
Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions |
title_fullStr |
Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions |
title_full_unstemmed |
Czerny-turner monochromator: Astigmatism in the classical and in the crossed beam dispositions |
title_sort |
czerny-turner monochromator: astigmatism in the classical and in the crossed beam dispositions |
url |
http://hdl.handle.net/20.500.12110/paper_1559128X_v25_n20_p3715_Simon |
work_keys_str_mv |
AT simonjm czernyturnermonochromatorastigmatismintheclassicalandinthecrossedbeamdispositions AT gilma czernyturnermonochromatorastigmatismintheclassicalandinthecrossedbeamdispositions AT fantinoan czernyturnermonochromatorastigmatismintheclassicalandinthecrossedbeamdispositions |
_version_ |
1807320800371408896 |